• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON & CO., (BD) 21 G X 0.75 IN. BD VACUTAINE® SAFETY-LOK¿ BLOOD COLLECTION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO., (BD) 21 G X 0.75 IN. BD VACUTAINE® SAFETY-LOK¿ BLOOD COLLECTION SET Back to Search Results
Catalog Number 367282
Device Problems Leak/Splash (1354); Improper Flow or Infusion (2954)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/19/2016
Event Type  malfunction  
Manufacturer Narrative
Results - bd had not received samples, but photos were provided by the customer facility for investigation.The photos were evaluated and the customer's indicated failure mode for leaking with the incident lot was not observed.A review of the manufacturing records was completed for the incident lot and no issues were identified.Conclusion: without a sample, an absolute root cause for this incident cannot be determined as bd was not able to duplicate or confirm the customer¿s indicated failure mode.
 
Event Description
It was reported that blood didn't flow into the collection set of the 21 g x 0.75 in.Bd vacutaine® safety-lok¿ blood collection set and leaked.There was air inside the sample and the tube stopped to fill.The same happened with another needle from the same lot: the venepuncture was done 3 times.No injury or medical intervention reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
21 G X 0.75 IN. BD VACUTAINE® SAFETY-LOK¿ BLOOD COLLECTION SET
Type of Device
BLOOD COLLECTION SET
Manufacturer (Section D)
BECTON, DICKINSON & CO., (BD)
1575 airport road
sumter SC 29153
Manufacturer (Section G)
BECTON, DICKINSON & CO., (BD)
1575 airport road
sumter SC 29153
Manufacturer Contact
brett wilko
9450 south state street
sandy, UT 84070
8015652845
MDR Report Key7059209
MDR Text Key93690475
Report Number1024879-2017-00908
Device Sequence Number1
Product Code JKA
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
K991088
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other,user facility
Reporter Occupation Other
Type of Report Initial
Report Date 11/07/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/27/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Phlebotomist
Device Expiration Date04/30/2018
Device Catalogue Number367282
Device Lot Number6112768
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received09/30/2016
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/03/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-