Model Number 10178573 |
Device Problem
Hole In Material (1293)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/21/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
The affected product has been received and the evaluation is pending.A follow up report will be submitted once the evaluation is completed.
|
|
Event Description
|
The customer reported that the patient was returned from operation room with the iv tubing infusing via gravity.The tubing was placed into the iv pump and the pump alarmed "air in line".The tubing was taken out of the pump and an empty syringe was used to draw back the air in the line and then placed back into the pump.The pump then alarmed 3 more times "air in line" before the nurse realized that there was a hole in the tubing.The tubing was disconnected from the patient and new tubing was hung without further complications.There was no report of patient harm.
|
|
Manufacturer Narrative
|
The customer¿s report of a hole in the tubing was confirmed.Visual inspection of the set noted that the silicone segment had a tear near the lower fitment.Examination under magnification showed no crush mark to the upper or lower fitment.Functional and pressure testing confirmed leaking from the silicone tubing near the lower fitment.The root cause of the air in the line was a tear in the silicone segment.The cause of the tear is unknown.
|
|
Search Alerts/Recalls
|