• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MDT KYPHON NEUCHATEL MFG KYPHX HV-R BONE CEMENT; CEMENT, BONE, VERTEBROPLASTY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MDT KYPHON NEUCHATEL MFG KYPHX HV-R BONE CEMENT; CEMENT, BONE, VERTEBROPLASTY Back to Search Results
Catalog Number C01A-J
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/01/2017
Event Type  malfunction  
Manufacturer Narrative
This product is not approved for sale in us but a similar device with catalog# c01a,510k # k041584 and udi (b)(4) is approved for sale in us.The device was not returned to manufacturer for evaluation therefore we are unable to determine the definitive cause of event.
 
Event Description
It was reported that intra-op, a broken container of liquid monomer was found.At that time, it smelled liquid monomer and liquid was found to adhere to the case.When liquid monomer was checked, the tip of the bottle was broken and it leaked.The event was noticed before it was transferred to clean area, and it was exchanged with new product.No patient complications were reported as a result of the event.
 
Manufacturer Narrative
Product analysis: the monomer vial is broken still inside its unopened package.The damage appears to have occurred during transit.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KYPHX HV-R BONE CEMENT
Type of Device
CEMENT, BONE, VERTEBROPLASTY
Manufacturer (Section D)
MDT KYPHON NEUCHATEL MFG
pierre-à-bot 97
neuchâtel 2000
CH  2000
Manufacturer (Section G)
MDT KYPHON NEUCHATEL MFG
pierre-à-bot 97
neuchâtel 2000
CH   2000
Manufacturer Contact
stacie ziemba
1800 pyramid place
memphis, TN 38132
9013963133
MDR Report Key7062480
MDR Text Key93563527
Report Number2953769-2017-00119
Device Sequence Number1
Product Code NDN
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
REFER H10
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/28/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2020
Device Catalogue NumberC01A-J
Device Lot NumberEL70022
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/28/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/05/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/28/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-