• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC COOK DOUBLE LUMEN EXTRA FIRM SOFT TIPPED AIRWAY EXCHANGE CATHETER; LRC CHANGER, TUBE, ENDOTRACHEAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC COOK DOUBLE LUMEN EXTRA FIRM SOFT TIPPED AIRWAY EXCHANGE CATHETER; LRC CHANGER, TUBE, ENDOTRACHEAL Back to Search Results
Model Number N/A
Device Problems Fracture (1260); Material Separation (1562)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/01/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The event is currently under investigation.A follow-up report will be submitted upon receipt of additional information or completion of the investigation.
 
Event Description
The international customer reported that, during the exchange of the endotracheal tube, the cook double lumen extra firm soft tipped airway exchange catheter became fractured and separated.The product malfunction reportedly occurred during contact with the patient; however, the customer confirmed that no additional procedures were required and no patient adverse events occurred as a result.The product is reportedly unavailable for return.
 
Manufacturer Narrative
Investigation ¿ evaluation: a review of the ifu (instructions for use) and quality control data was conducted during the investigation.The complaint device was not returned, so physical evaluation could not be performed.Additionally, no lot number was provided; therefore a device history record review and complaint lot search cannot be completed at this time.However, a review of the quality control data and product labeling was performed.There is no indication that a design or process related failure mode contributed to this event.Current controls for manufacturing are in place to assure functionality and device integrity prior to shipping.Review of production and quality documentation did not observe any specific issues with current manufacturing or quality controls that may have contributed to this incident.The device is shipped with instructions for use, which states the proper warnings, precautions, and instructions for use.Based on the provided information a definitive root cause cannot be established or reported at this time.Per the quality engineering risk assessment no further action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COOK DOUBLE LUMEN EXTRA FIRM SOFT TIPPED AIRWAY EXCHANGE CATHETER
Type of Device
LRC CHANGER, TUBE, ENDOTRACHEAL
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
larry pool
750 daniels way
bloomington, IN 47404
8128294891
MDR Report Key7063080
MDR Text Key93326025
Report Number1820334-2017-04195
Device Sequence Number1
Product Code LRC
Combination Product (y/n)N
Reporter Country CodeKR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,o
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue NumberC-CAE-11.0-100-DLT-EF-ST
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/02/2017
Initial Date FDA Received11/28/2017
Supplement Dates Manufacturer Received02/21/2018
Supplement Dates FDA Received03/21/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-