Model Number B-4800
Device Problem Free or Unrestricted Flow (2945)
Patient Problem Vomiting (2144)
Event Date 10/11/2017
Event Type  malfunction  
Manufacturer Narrative
The reporter of the event was asked to return the product for analysis.To date, apollo has not received the device.Device labeling addresses the reported events as follows: precautions: antiemetics, antispasmodic, and anticholinergic drugs may be prescribed to lessen the early placement symptoms such as nausea, vomiting, and abdominal pain.Patients will need to immediately contact their physician for any severe or unusual symptoms.Placement of the balloon within the stomach produces an expected and predictable reaction characterized most commonly by a feeling of heaviness in the abdomen, nausea and vomiting, gastroesophageal reflux, belching, esophagitis, heartburn, diarrhea and, at times, abdominal, back or epigastric pain and cramping.Food digestion may be slowed during this adjustment period.These symptoms can be treated with antiemetic, antispasmodic, and anticholinergic medications.Typically the stomach acclimates to the presence of the device within the first 2 weeks.In order to prevent or ameliorate the symptoms most frequently experienced during the adjustment period, it is recommended that the physician use proton pump inhibitors (ppis), antiemetics, antispasmodics, and anticholinergic medications prophylactically (before orbera® placement).Patients should be advised to immediately contact their physician for any unusually severe or worsening symptoms.The physiological response of the patient to the presence of orbera® may vary depending upon the patient's general condition and the level and type of activity.The types and frequency of administration of drugs or diet supplements and the overall diet of the patient may also affect the response.Each patient must be monitored closely during the entire term of treatment in order to detect the development of possible complications.Each patient should be instructed regarding symptoms of deflation, gastrointestinal obstruction, ulceration and other complications which might occur, and should be advised to contact his/her physician immediately upon the onset of such symptoms.Patients need to be evaluated and the device removed at or within 6 months of placement.Clinical data does not exist to support use of an individual orbera® beyond 6 months.Adverse events: it is important to discuss all possible complications and adverse events with your patient.Complications that may result from the use of this product include the risks associated with the medications and methods utilized in the endoscopic procedure, the risks associated with any endoscopic procedure, the risks associated with the orbera intragastric balloon specifically, and the risks associated with the patient's degree of intolerance to a foreign object placed in the stomach.Possible complications - possible complications of the use of orbera® include: - gastric discomfort, feelings of nausea and vomiting following balloon placement as the digestive system adjusts to the presence of the balloon.-continuing nausea and vomiting.This could result from direct irritation of the lining of the stomach or as a result of the balloon blocking the outlet of the stomach.It is even theoretically possible that the balloon could prevent vomiting (not nausea or retching) by blocking the inlet to the stomach from the esophagus.- spontaneous over inflation warnings: spontaneous over inflation of an indwelling balloon has been reported in patients with orbera®.Symptoms of balloon over-inflation include intense abdominal pain, swelling of the abdomen (abdominal distension) with or without discomfort, difficulty breathing, and/or vomiting.Patients experiencing any of these symptoms should be counseled to seek immediate care.
 
Event Description
Reported as: a patient with the orbera intragastric balloon had experienced vomiting a week after insertion.The physician removed the device early due to spontaneous inflation.
 
Manufacturer Narrative
Device evaluation summary: the device was returned to apollo, and a visual inspection was performed.White and yellow particles were noted on the outer surface of the balloon shell.As the balloon was not received with the fill tube, a sample fill tube was used for device testing.A valve test was performed, and when di water was injected into the valve, the flow of fluid was continuous and unobstructed.An air leak test was performed, and the balloon was noted to be leaking from a single opening on the radius of the shell.Under microscopic analysis, the opening in the shell was noted to have striated edges, consistent with surgical damage from a device removal tool.Blue particulate matter was observed in the balloon valve.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORBERA INTRAGASTRIC BALLOON SYSTEM
Type of Device
INTRAGASTRIC BALLOON
Manufacturer (Section D)
APOLLO ENDOSURGERY, INC
1120 s. capital of texas hwy
bldg. 1, ste. 300
austin TX 78746
Manufacturer (Section G)
APOLLO ENDOSURGERY COSTA RICA, SRL
coyol free zone
building b 13.3
alajuela, costa rica
CS  
Manufacturer Contact
laura lebouef
1120 s. capital of texas hwy
bldg. 1, ste. 300
austin, TX 78746
5122795141
MDR Report Key7068392
MDR Text Key93694832
Report Number3006722112-2017-00397
Device Sequence Number1
Product Code LTI
UDI-Device Identifier10811955020619
UDI-Public10811955020619
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P140008
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date04/26/2019
Device Model NumberB-4800
Device Catalogue NumberB-4800
Device Lot NumberAF00700
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/04/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/02/2017
Initial Date FDA Received11/29/2017
Supplement Dates Manufacturer Received02/21/2018
Supplement Dates FDA Received03/20/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/26/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age42 YR
-
-