• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC CRYOCATH LP FLEXCATH ADVANCE STEERABLE SHEATH; CATHETER, STEERABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC CRYOCATH LP FLEXCATH ADVANCE STEERABLE SHEATH; CATHETER, STEERABLE Back to Search Results
Model Number 4FC12
Device Problem Air Leak (1008)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/14/2017
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during a cryo ablation procedure, air ingress was observed during aspiration.The sheath was replaced with resolve, and the case was completed with cryo.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Event summary: the patient data files showed eleven injections were performed with catheter 2af284/56483-3 and showed temperature was not able to go below -35 degrees celsius at the first and second injections.Upon visual inspection of flexcath sheath 4fc12/69967-075, results showed the device was intact with no apparent issues.Functional testing of the sheath confirmed the hemostatic valve was leaking.Air bubbles were observed through the valve.It was suspected that the valve disk was torn.In conclusion, the reported air ingress issue has been confirmed through testing.The sheath failed the returned product inspection due to a leaking hemostatic valve.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXCATH ADVANCE STEERABLE SHEATH
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA  H9R 5Z8
Manufacturer (Section G)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA   H9R 5Z8
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key7076573
MDR Text Key93726890
Report Number3002648230-2017-00664
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K123591
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/21/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/08/2019
Device Model Number4FC12
Device Catalogue Number4FC12
Device Lot Number69967
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/14/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/14/2017
Initial Date FDA Received12/01/2017
Supplement Dates Manufacturer Received12/06/2017
Supplement Dates FDA Received12/21/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/08/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-