Catalog Number 03260763160 |
Device Problem
High Test Results (2457)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 11/15/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
The customer received false positive nitrate results for three patient samples and a false positive leukocyte result for two of the patient urine samples.Of the data provided, only the results for nitrate were discrepant.Information was provided that one sample was sent to the laboratory for repeat resting and the result was negative.The laboratory results were considered to be correct.There was no allegation of an adverse event.The customer used urisys 1100 analyzer serial number (b)(4).
|
|
Manufacturer Narrative
|
Additional information was received that the laboratory analyzer used was clinitek 500.
|
|
Manufacturer Narrative
|
Retention material of lot #18697700 was measured on a (b)(4) with a native urine sample, a nitrite- sample, and a leucocyte- dilution series and was visually checked according to the testing plan.The results fulfilled the requirements.No false negative or false positive results were observed.
|
|
Manufacturer Narrative
|
One test strip container with 44 out of 100 test strips was received for investigation.The test strips showed a light dusty shade.The urisys 1100 analyzer was also received but was missing the test strip tray.The analyzer showed no signs of damage.Retention material of lot #18697700 and the customer material of lot #186 977 01 was tested on a retention urisys1100 analyzer and on the customer's analyzer urisys 1100 native urine, a nitrite- and a leucocyte-dilution-series.The results of the retention material measurements fulfill the requirements.No false negative or false positive results were observed.The results of the customer material measurements did not fulfill the requirements.False positive results for nitrate and leukocyte were received.The customer analyzer worked properly.The investigation was unable to find a definitive root cause.It appeared the customer strips had not been properly stored (tube not closed properly) and therefore the strips generate higher (false positive) results in comparison to retention material.
|
|
Search Alerts/Recalls
|