• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON & CO., (BD) 13X75 MM 3.5 ML BD VACUTAINER® PLUS PLASTIC SST TUBE. GOLD BD HEMOGARD¿ CLOSURE; BLOOD COLLECTION TUBE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO., (BD) 13X75 MM 3.5 ML BD VACUTAINER® PLUS PLASTIC SST TUBE. GOLD BD HEMOGARD¿ CLOSURE; BLOOD COLLECTION TUBE Back to Search Results
Catalog Number 367983
Device Problem Detachment Of Device Component (1104)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/06/2015
Event Type  malfunction  
Manufacturer Narrative
Results: bd had not received samples or photos from the customer facility for evaluation; therefore, the investigation was limited.A review of the manufacturing records was completed for the incident lot and no issues were identified.Conclusion: without a sample, an absolute root cause for this incident cannot be determined as bd was not able to duplicate or confirm the customer¿s indicated failure mode.
 
Event Description
It was reported that while transferring blood from a 10 cc syringe with the maximus mbc6008 into a 13x75 mm 3.5 ml bd vacutainer® plus plastic sst tube.Gold bd hemogard¿ closure, the hemogard top came off while removing the syringe.While this occurred blood splashed onto the nurse.No serious injury or medical intervention reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
13X75 MM 3.5 ML BD VACUTAINER® PLUS PLASTIC SST TUBE. GOLD BD HEMOGARD¿ CLOSURE
Type of Device
BLOOD COLLECTION TUBE
Manufacturer (Section D)
BECTON, DICKINSON & CO., (BD)
1575 airport road
sumter SC 29153
Manufacturer (Section G)
BECTON, DICKINSON & CO., (BD)
1575 airport road
sumter SC 29153
Manufacturer Contact
brett wilko
9450 south state street
sandy, UT 84070
8015652845
MDR Report Key7080944
MDR Text Key94671843
Report Number1024879-2017-00276
Device Sequence Number1
Product Code JKA
UDI-Device Identifier00382903679836
UDI-Public00382903679836
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K023075
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other,use
Reporter Occupation Other
Type of Report Initial
Report Date 11/24/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Nurse
Device Expiration Date05/31/2016
Device Catalogue Number367983
Device Lot Number5148534
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/13/2015
Initial Date FDA Received12/04/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/28/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-