Model Number N/A
Device Problem Difficult to Insert (1316)
Patient Problem No Code Available (3191)
Event Date 11/08/2017
Event Type  Injury  
Manufacturer Narrative
(b)(4).Concomitant medical products: 010000857, g7 neutral e1 liner 36mm e, 6134049.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001825034 - 2017 - 10930.
 
Event Description
It was reported that during the primary hip surgery the liner would not fit into the shell.A second liner was used, and the same issue occured.The surgery was delayed by more than thirty minutes.Attempts have been made and additional information on the reported event is unavailable.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.Visual inspection identified several nicks around the rim and scuffs to the articulating surface.The locking flange and scallops have a few nicks and indents from the impaction and removal.Dhr was reviewed and no discrepancies relevant to the reported event were found.Review of the complaint history determined that no further action is required.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 NEUTRAL E1 LINER 36MM E
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key7086097
MDR Text Key93814468
Report Number0001825034-2017-10749
Device Sequence Number1
Product Code PBI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK121874
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 04/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date09/04/2022
Device Model NumberN/A
Device Catalogue Number010000857
Device Lot Number6126149
Other Device ID Number(01) 00880304526426
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/02/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/08/2017
Initial Date FDA Received12/06/2017
Supplement Dates Manufacturer Received04/04/2018
04/04/2018
Supplement Dates FDA Received04/04/2018
04/04/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/06/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age66 YR
-
-