Model Number 479688
Device Problems Device Operates Differently Than Expected (2913); Device Dislodged or Dislocated (2923)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/12/2017
Event Type  malfunction  
Event Description
It was reported that during the implant procedure, the left ventricular (lv) lead pulled out of the coronary sinus on three occasions.A stable position was obtained and the lead seemed secure.It was noted that the lead was selected due to difficult anatomy.During the night, the lv lead dislodged.It was noted that the lead did not seem to ¿screw¿ into the vessel.The lv lead was explanted and replaced with a different lead to offer better stability.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTAIN STABILITY
Type of Device
DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key7094296
MDR Text Key94252389
Report Number2649622-2017-15280
Device Sequence Number1
Product Code OJX
Combination Product (y/n)N
Reporter Country CodeSF
PMA/PMN Number
P080006
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 12/07/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/07/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/16/2019
Device Model Number479688
Device Catalogue Number479688
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/12/2017
Date Device Manufactured01/24/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age70 YR
Patient Weight51
-
-