• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC SYNCARDIA COMPANION 2 DRIVER; EXTERNAL PNEUMATIC DRIVER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC SYNCARDIA COMPANION 2 DRIVER; EXTERNAL PNEUMATIC DRIVER Back to Search Results
Model Number 397002-001
Device Problems Self-Activation or Keying (1557); Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/04/2017
Event Type  malfunction  
Manufacturer Narrative
This alleged failure mode poses a low risk to the patient because it did not prevent the companion 2 driver from performing its life-sustaining functions.The companion 2 driver will be returned to syncardia for evaluation.The results of the evaluation will be provided in a follow-up mdr.(b)(4).
 
Event Description
The customer, a syncardia certified hospital, reported that the companion 2 driver compressors cycled on and off while the driver was connected to wall air and supporting a patient.The customer also reported that the patient was subsequently switched to the backup companion 2 driver.There was no reported adverse patient impact.
 
Manufacturer Narrative
The companion 2 driver was returned to syncardia for evaluation.The customer-reported issue of the driver's compressor cycling on and off was confirmed.Investigation testing determined the root cause to be a malfunction of the manual pressure regulator.Syncardia has a corrective and preventive action (capa) to address the issue of compressor cycling.Syncardia has completed its evaluation of this complaint and is closing this file.(b)(4) follow-up report 1.
 
Event Description
The customer, a syncardia certified hospital, reported that the companion 2 driver compressors cycled on and off while the driver was connected to wall air and supporting a patient.The customer also reported that the patient was subsequently switched to the backup companion 2 driver.There was no reported adverse patient impact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA COMPANION 2 DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer Contact
michael garippa
1992 e. silverlake road
tucson, AZ 85713
5205451234
MDR Report Key7099499
MDR Text Key95757393
Report Number3003761017-2017-00250
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003107
UDI-Public(01)00858000003107
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Biomedical Engineer
Remedial Action Replace
Type of Report Initial,Followup
Report Date 12/04/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number397002-001
Device Catalogue Number397002-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/07/2017
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/04/2017
Initial Date FDA Received12/08/2017
Supplement Dates Manufacturer Received12/21/2017
Supplement Dates FDA Received01/02/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/13/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age28 YR
-
-