Model Number H74939208181650 |
Device Problem
Device Operates Differently Than Expected (2913)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 11/27/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Age at time of event: 18 years or older.(b)(4).
|
|
Event Description
|
It was reported that the device outer shaft rotated.The chronic total occlusion was located in the mildly tortuous and moderately calcified superior femoral artery (sfa).A truepath¿ cto device was selected for use.During preparation, it was noted that the outer shaft rotated instead of the tip.The procedure was completed with another of the same device.No patient complications were reported.
|
|
Manufacturer Narrative
|
Device evaluated by manufacturer: the device was returned for analysis.Returned product consisted of a truepath device.The control unit, motor housing and working length and tip/tip housing were microscopically examined.There was saline along on the shaft and tip of the device.There was no damage or irregularities identified during analysis.Functional testing performed and the device turned on and the normal audible noise was heard, meanwhile the device vibrated and the burr tip rotated with no alert light or issues.When pressure was applied and released the alert lights and audible noise worked properly.Inspection of the device presented no damage or irregularities.Device analysis determined the condition of the returned device was not consistent with the reported information.The investigation conclusion is not confirmed - returned as there was no evidence of either the alleged issue(s) or any defect which could have contributed to the event.(b)(4).
|
|
Event Description
|
It was reported that the device outer shaft rotated.The chronic total occlusion was located in the mildly tortuous and moderately calcified superior femoral artery (sfa).A truepath¿ cto device was selected for use.During preparation, it was noted that the outer shaft rotated instead of the tip.The procedure was completed with another of the same device.No patient complications were reported.
|
|
Search Alerts/Recalls
|