Catalog Number 7243
Device Problems Failure To Adhere Or Bond (1031); Detachment Of Device Component (1104); Product Quality Problem (1506)
Patient Problem No Information (3190)
Event Date 01/26/2017
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event Description
It was reported that the sheaths had an insufficient amount of adhesive on them which caused them to come off during the night.During sample evaluation, lots were found to have strong adhesion.
 
Manufacturer Narrative
Received 3 unopened spirit mecs.During the investigation of a separate complaint it was found that one of the mec's returned had strong adhesive.For that reason, that sample was used for this investigation.The samples were visually inspected for holes or damage on the package, legibility of printed numbers and letters, expiration date, correct impression, print color, missing parts of printing, position of printing, damaged component, and foreign matter.The sample was functionally tested under the male external catheter adhesive peel strength test.The tested sample was not within specification - the acceptable peel strength must be in a range from 0.60 to 2.10 lbf for a mec 6030 po 01 s 25 mm and sample results were: sample 1: 2.53 lbf the device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: "fitting the sheath is a strapless penile sheath: it has an adhesive coating inside to ensure a safe and simple procedure for fitting.If necessary trim pubic hair.Clean and dry pubic area.Remove sheath from its wrapping.Place rolled end over the end of the penis, leaving a small space between the end of the penis and the cup of the sheath.If uncircumcised, no not retract the foreskin but allow it to remain over the glans.Avoid contact between the adhesive and the glans.Slowly unroll the sheath along the shaft of the penis, then gently squeeze the sheath around the penis to ensure even adhesion.Try to avoid leaving a rolled 'collar' of sheath around the base of the penis.Finally, connect the urine collection bag, always check that the connections are secure before use.Wear time will vary from user to user.It is recommended that a sheath be changed every 24 hours for reasons of hygiene.To remove: the sheath may be removed by gently rolling it off the penis.Avoid simply pulling the sheath off.Removing the sheath whilst having a bath or shower may increase comfort.Do not use on irritated or compromised skin.Discontinue use if irritation occurs." (b)(4).
 
Event Description
It was reported that the sheaths had an insufficient amount of adhesive on them which caused them to come off during the night.During sample evaluation, lots were found to have strong adhesion.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROCHESTER CLEAR ADVANTAGE SPIRIT HYDRO SHEATH STYLE 2 24MM BX30
Type of Device
MEC
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer Contact
amy gravley
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key7107622
MDR Text Key95717284
Report Number1018233-2017-06325
Device Sequence Number1
Product Code NNX
UDI-Device Identifier00801741071638
UDI-Public(01)00801741071638
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,other
Reporter Occupation Patient
Type of Report Initial,Followup
Report Date 12/21/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/12/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date09/30/2021
Device Catalogue Number7243
Device Lot NumberJUAX0695
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/20/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received12/12/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/27/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-