The returned device matches the upn and lot number provided by the customer visual inspection showed that there is a kink located approximately 15cm from the distal tip proximal to ring 3.There is dried body fluid and saline in the irrigation ports in the tip; the ring seals are intact.Dried body fluid is on the handle and along the left seam of the handle, along the hole plug, on both sides of the distal end of the handle, and around the strain relief.The seams of the handle appear to be intact.The butt bond seal is compromised.There is body fluid present along the seal.Continuity checks revealed no electrical shorts as checked manually using a multi-meter and breakout box.Functional testing showed the steering knob and the tension control knob functioned properly on both lock and unlock positions.Temperature testing showed that the catheter met expected values.Device was dissected and dried blood is throughout the entire length of the handle interior.There is also a large amount of the dried body fluid throughout the interior of the shaft.X-rays of the sensor revealed that one of the connecting wires between the sensor wire and the sensor body is fractured.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is design specification as the device problem was traced back to the design specifications.(b)(4).
|