Model Number H7493925138250
Device Problems Bent (1059); Device Damaged Prior to Use (2284)
Patient Problem No Patient Involvement (2645)
Event Date 11/29/2017
Event Type  malfunction  
Manufacturer Narrative
Device is a combination product.(b)(4).Device evaluated by mfr.: the device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.(b)(4).
 
Event Description
It was reported that stent damage occurred.During unpacking of a 38 x 2.50 promus premier¿ drug-eluting stent, it was noted that the stent strut was lifted up.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.
 
Manufacturer Narrative
Device evaluated by mfr.: promus premier ous mr 38 x 2.50mm stent delivery system was returned for analysis.A visual and microscopic examination of the crimped stent identified distal stent damage.Damage was noted to the two most distal stent strut rows; struts had been pulled distally.The undamaged crimped stent outer diameter was measured and was within max crimped stent profile measurement.The balloon cones were reviewed and damage was noted; a tear was found in the distal balloon cone immediately distal of the distal stent struts.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and tactile examination of the hypotube found no issues.A visual and tactile examination of the shaft polymer extrusion revealed no issues.A visual and microscopic examination of the tip revealed no issues.No other issues were identified during analysis.The investigation conclusion is handling damage as the complaint was caused by handling of the device or portion of the device without direct patient contact.(b)(4).
 
Event Description
It was reported that stent damage occurred.During unpacking of a 38 x 2.50 promus premier¿ drug-eluting stent, it was noted that the stent strut was lifted up.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROMUS PREMIER¿
Type of Device
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
Manufacturer (Section G)
BOSTON SCIENTIFIC - GALWAY
Manufacturer Contact
sonali arangil
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key7110542
MDR Text Key94967265
Report Number2134265-2017-12562
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)N
Reporter Country CodeCN
PMA/PMN Number
P110010
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/04/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/12/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/08/2019
Device Model NumberH7493925138250
Device Catalogue Number39251-3825
Device Lot Number0020409901
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/21/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/11/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/24/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-