Complaint sample was evaluated and the reported event was confirmed.Visual examination showed there is significant wear to the etch marks on the instrument guide, rendering the guide unreadable dhr was reviewed and no discrepancies relevant to the reported event were found.Review of the complaint history determined that no further action is required as no were trends identified.Based on the amount of cosmetic damage and the device's service life, the cause of this event is determined to be expected wear.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|