|
Catalog Number RBY4C0860 |
Device Problems
Bent (1059); Device Handling Problem (3265)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 11/13/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Results: the pet lock was intact on the proximal end of the second ruby coil evaluated.The pusher assembly was kinked approximately 5.0 cm from the proximal end.The stretch resistant wire (sr wire) was fractured and the proximal constraint sphere was inside the distal detachment tip (ddt).The embolization was detached from its pusher assembly and inside its introducer sheath.Conclusions: evaluation of the first returned ruby coil revealed that the pusher assembly was fractured.This type of likely occurred from the pusher assembly being accidentally kinked as mentioned in the complaint.Then further manipulation of the kinked pusher likely resulted in the pusher assembly fracturing as the kink was straightened.Further evaluation of the returned device revealed that the embolization coil was detached from its pusher assembly.The separation of the two fractured pusher assembly segments allowed the pull wire to be retracted out of its ddt caused the embolization coil to detach.Evaluation of the second returned ruby coil revealed that its pusher assemblies was kinked.This damage was likely accidental as mentioned in the complaint.Further evaluation revealed that the sr wire was fractured and the proximal constraint sphere was inside the distal detachment tip (ddt).This type of damage likely occurred due to forceful retraction against resistance, while re-sheathing the embolization coil as it was returned halfway re-sheathed inside its introducer sheath.The non-penumbra microcatheter mentioned in the complaint was not returned for evaluation.Penumbra coils are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report number: 3005168196-2017-02195.
|
|
Event Description
|
The patient was undergoing a coil embolization procedure in the splenic artery using ruby coils.During the procedure, the physician successfully deployed and detached two ruby coils into the target vessel.However, while attempting to advance two new ruby coils through a non-penumbra microcatheter, the scrub technologist inadvertently bent the back end of the ruby coils; therefore, both ruby coils were removed.The procedure was then completed using a new ruby coil and the same microcatheter.There was no report of an adverse effect to the patient.
|
|
Search Alerts/Recalls
|
|
|