• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LITTLETON) O-ARM 1000 IMAGING SYSTEM; IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LITTLETON) O-ARM 1000 IMAGING SYSTEM; IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE Back to Search Results
Model Number BI-700-00027-120
Device Problems Computer Software Problem (1112); Application Interface Becomes Non-Functional Or Program Exits Abnormally (1138); Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/16/2017
Event Type  malfunction  
Manufacturer Narrative
Patient information was unavailable from the site.No parts have returned to the manufacturer for evaluation.
 
Event Description
A site representative reported that, while in an electrode and probe placement procedure, the imaging system was booting on its own.The site reported that the system rebooted twice on its own during the procedure.The second time the system remained in stand alone mode for 10 minutes.When the site rebooted the entire system again, a blue screen was displayed.In addition, the handswitch is not working properly.The first button does not respond when pressed.There was a reported delay to the procedure of less than 1 hour due to this issue and no impact on the patient outcome.
 
Manufacturer Narrative
A medtronic representative went to the site to test the equipment.The imaging system passed the system checkout and was found to be fully functional.No hardware parts were replaced.The reported event could not be duplicated by medtronic personnel.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-ARM 1000 IMAGING SYSTEM
Type of Device
IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer Contact
amos jarrette
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
7208903200
MDR Report Key7114594
MDR Text Key95056897
Report Number1723170-2017-05124
Device Sequence Number1
Product Code OXO
Combination Product (y/n)N
Reporter Country CodeES
PMA/PMN Number
K050996
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 04/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Radiologic Technologist
Device Model NumberBI-700-00027-120
Device Catalogue Number9732719
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/16/2017
Initial Date FDA Received12/13/2017
Supplement Dates Manufacturer Received11/22/2017
Supplement Dates FDA Received04/18/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/15/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-