• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET FRANCE S.A.R.L. REFOBACIN BONE CEMENT R 2X40G; BONE CEMENT, ANTIBIOTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET FRANCE S.A.R.L. REFOBACIN BONE CEMENT R 2X40G; BONE CEMENT, ANTIBIOTIC Back to Search Results
Model Number N/A
Device Problem Unsealed Device Packaging (1444)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/01/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).(b)(6).The product within this report is a combination product.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Not returned to manufacturer.
 
Event Description
It was reported the inner sterile packaging was compromised and allowed the contents to leak out.No further information is available at this time.
 
Manufacturer Narrative
(b)(4).Complaint device was evaluated and the reported event was confirmed.Evaluation of returned device showed that the supplier sealing was opened.The manufacturer sealing was intact.Device history record (dhr) was reviewed and no discrepancies were found.Investigation concluded that the reported event was due to supplier packaging issue.Corrective action has been initiated to address the reported issue.If any further information is found which would change or alter any conclusions, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REFOBACIN BONE CEMENT R 2X40G
Type of Device
BONE CEMENT, ANTIBIOTIC
Manufacturer (Section D)
BIOMET FRANCE S.A.R.L.
plateau de lautagne bp75
valence cedex 26903
FR  26903
Manufacturer (Section G)
BIOMET FRANCE S.A.R.L.
plateau de lautagne bp75
valence cedex 26903
FR   26903
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key7120770
MDR Text Key95745497
Report Number3006946279-2017-00284
Device Sequence Number1
Product Code MBB
Combination Product (y/n)N
Reporter Country CodeMY
PMA/PMN Number
PK150850
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/15/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2018
Device Model NumberN/A
Device Catalogue Number4003940002
Device Lot NumberA519AA2801
Other Device ID NumberSEE H10 NARRATIVE
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received11/20/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/20/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Age71 YR
-
-