• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL HUDSON CIRCUIT,(22MM),KRATON,DUAL HTD LIMBS; HUMIDIFIER, RESPIRATORY GAS, (

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL HUDSON CIRCUIT,(22MM),KRATON,DUAL HTD LIMBS; HUMIDIFIER, RESPIRATORY GAS, ( Back to Search Results
Catalog Number 790-32
Device Problem Leak/Splash (1354)
Patient Problem No Patient Involvement (2645)
Event Date 12/06/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).A visual, dimensional, and functional inspection of the device involved in the complaint could not be conducted since the device was not returned to the manufacturer at the time of this report.Device history record review shows that the product was assembled and inspected according to our specifications.Customer complaint cannot be confirmed based only on the information provided.If the device sample becomes available at a later date this report will be updated as applicable.
 
Event Description
Customer complaint alleges the device leaked during the short service test on the ventilator.Customer reports there was no patient involvement.
 
Manufacturer Narrative
(b)(4).The sample was returned for evaluation.A visual exam was performed and no defects were observed.In addition, no dimensional issues were identified.The sample met the current manufacturing specification.Functional testing was performed and the sample failed the leak test.The sample was then submerged in water and bubbles were observed, confirming there was a leak.The customer complaint is confirmed based on the functional inspection.The sample did not pass the leak test.A non-conformance report was initiated to complete and document the investigation and to provide corrective actions.
 
Event Description
Customer complaint alleges the device leaked during the short service test on the ventilator.Customer reports there was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HUDSON CIRCUIT,(22MM),KRATON,DUAL HTD LIMBS
Type of Device
HUMIDIFIER, RESPIRATORY GAS, (
Manufacturer (Section D)
TELEFLEX MEDICAL
research triangle park NC
Manufacturer (Section G)
TELEFLEX MEDICAL
parque industrial finsa
nuevo laredo 88275
MX   88275
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
9194334854
MDR Report Key7122379
MDR Text Key95059529
Report Number3004365956-2017-00455
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K031383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/06/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number790-32
Device Lot Number74E1700107
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/15/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/06/2017
Initial Date FDA Received12/15/2017
Supplement Dates Manufacturer Received01/31/2018
Supplement Dates FDA Received01/31/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/01/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-