Model Number SC-2316-50 |
Device Problems
Fracture (1260); High impedance (1291); Kinked (1339)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 11/29/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Additional suspect medical device components involved in the event: model #: sc-2316-50 serial #: (b)(4), description: infinion 1x16 percutaneous lead kit-50cm model #: sc-3400-30 serial #: (b)(4), description: infinion splitter 2x8 kit (30 cm).
|
|
Event Description
|
A report was received that the patients contacts were showing high impedances.The patient underwent a full scs system replacement procedure.
|
|
Manufacturer Narrative
|
Additional information was received that the physician had explanted and replaced all of the patients old scs system.Sc-1132 (sn: (b)(4)) device evaluation indicated that the ipg passed all tests performed.Sc-2316-50 (sn: (b)(4)) device evaluation indicated that the complaint was confirmed.Visual inspection revealed that the lead body was kinked right after the retention sleeve in the proximal end.All 16 cables were fractured at the damaged section of the lead and it resulted in high impedances.No cables were exposed at the site.Sc-2316-50 (sn: (b)(4)) device evaluation indicated that the complaint was confirmed.14 cables were fractured on the lead body at the kinked section where the clik-anchor was positioned.Fracture location is 1 cm from the clik anchor set screw mark.The fractured cables resulted in high impedances.No cables were exposed.Sc-3400-30 (sn: (b)(4)) device evaluation indicated that the complaint was confirmed.E#14 cable was broken at the weld nugget of the contact spring inside the boot.No cables were exposed at the site.Sc-3400-30 (sn: (b)(4)) device evaluation indicated that the device passed all tests performed.Sc-1318: device evaluation indicated that the eyelets were torn, and silicone material from one of the eyelets was missing.
|
|
Event Description
|
A report was received that the patients contacts were showing high impedances.The patient underwent a full scs system replacement procedure.
|
|
Manufacturer Narrative
|
Correction to mdr fu#1 in field.Should have been; additional information was received that the physician had explanted and replaced all of the patients old scs system.Sc-1132 (sn: (b)(4)) device evaluation indicated that the ipg passed all tests performed.Sc-2316-50 (sn: (b)(4)) device evaluation indicated that the complaint was confirmed.Visual inspection revealed that the lead body was kinked right after the retention sleeve in the proximal end.All 16 cables were fractured at the damaged section of the lead and it resulted in high impedances.No cables were exposed at the site.Sc-2316-50 (sn: (b)(4)) device evaluation indicated that the complaint was confirmed.14 cables were fractured on the lead body at the kinked section where the clik-anchor was positioned.Fracture location is 1 cm from the clik anchor set screw mark.The fractured cables resulted in high impedances.No cables were exposed.Sc-3400-30 (sn: (b)(4)) device evaluation indicated that the complaint was confirmed.E#14 cable was broken at the weld nugget of the contact spring inside the boot.No cables were exposed at the site.Sc-3400-30 (sn: (b)(4)) device evaluation indicated that the device passed all tests performed.Sc-4316: device evaluation indicated that the eyelets were torn, and silicone material from one of the eyelets was missing.
|
|
Event Description
|
A report was received that the patients contacts were showing high impedances.The patient underwent a full scs system replacement procedure.
|
|
Manufacturer Narrative
|
Correction to mdr fu #2 mdr in field - should have been (b)(6) 2018.
|
|
Event Description
|
A report was received that the patients contacts were showing high impedances.The patient underwent a full scs system replacement procedure.
|
|
Search Alerts/Recalls
|