ACCLARENT, INC. RELIEVA SPIN PLUS KIT, 3 GUIDES, 6MM, 5PK; INSTRUMENT, ENT MANUAL SURGICAL Back to Search Results
Catalog Number RSP0616MFS
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/16/2017
Event Type  malfunction  
Manufacturer Narrative
The patient identifier is not known.The patient¿s age is not known.The patient's weight is not known.The lot number of the device is not known.The full udi # information is unavailable since the lot number is unavailable.The lot history record review could not be performed as lot number for the acclarent relieva spinplus device was unavailable.If additional information is received regarding this event, a supplemental report will be filed.Manufacturer reference number: (b)(4).
 
Event Description
Acclarent was informed on (b)(6) 2017, for an event that occurred on (b)(6) 2017 involving a relieva spinplus kit, 3 guides, 6mm, 5pk (lot# not available).During the primary hybrid balloon sinuplasty procedure, the physician noticed that the blue tip from the balloon catheter had separated from the balloon and was in the sinus cavity.The blue tip was retrieved with forceps.There was no report of any patient consequence and the procedure was not extended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RELIEVA SPIN PLUS KIT, 3 GUIDES, 6MM, 5PK
Type of Device
INSTRUMENT, ENT MANUAL SURGICAL
Manufacturer (Section D)
ACCLARENT, INC.
33 technology drive
irvine 92618
Manufacturer (Section G)
ACCLARENT, INC.
33 technology drive
irvine 92618
Manufacturer Contact
joaquin kurz
33 technology drive
irvine 92618
949789-383
MDR Report Key7132458
MDR Text Key95641212
Report Number3005172759-2017-00057
Device Sequence Number1
Product Code LRC
Combination Product (y/n)N
PMA/PMN Number
K143541
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Health Professional
Type of Report Initial
Report Date 11/29/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberRSP0616MFS
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/29/2017
Initial Date FDA Received12/19/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-