It was reported that the embolization coil did not detach with two v-grip controllers.During removal, the coil unexpectedly detached in the microcatheter.The coil was then advanced and implanted in the aneurysm with the delivery wire.There was no reported intervention, patient injury, or health damage.
|
The pusher, introducer, and v-grip controller were returned for analysis.During the investigation, the strain relief was placed on the distal heater coil section, and it appeared burnt, with an irregular non-uniform tip shape.The pusher was noted to be broken at the pet transition.The remainder of the pusher was noted to be within specification, with no kinks or damage.The monofilament within the body coil was analyzed to check on its shape and rebound length; the monofilament appeared to have a stretched tail.Based on the investigation, the reported detachment was confirmed but the analysis showed that the pet was burnt, which is an indication that detachment was attempted, although the coil retaining monofilament appeared to have tensile overload as if an incomplete detachment may have occurred.The returned v-grip was found to be performing to requirements.
|