• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LITTLETON) O-ARM 1000 IMAGING SYSTEM 3RD EDITION; IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LITTLETON) O-ARM 1000 IMAGING SYSTEM 3RD EDITION; IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE Back to Search Results
Model Number BI-700-00028-120
Device Problem Mechanical Problem (1384)
Patient Problem No Patient Involvement (2645)
Event Date 11/27/2017
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.A medtronic representative went to site to test the equipment.Representative reported that a x-stage cover was replaced and invalidate home was performed.Issue resolved.A system checkout was performed and the hardware, software, and instruments passed the system checkout.The system was found to be fully functional.The suspect x-stage cover has been received by the manufacturer for evaluation.The reported issue could not be confirmed as the returned cover failed visual inspection with physical damage.X-stage cover had multiple scratches and dents around the docking pin holes.
 
Event Description
A medtronic representative reported that while outside of procedure, the imaging system would not dock.Invalidate home was performed but the issue persisted.There was no patient present at the time of the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-ARM 1000 IMAGING SYSTEM 3RD EDITION
Type of Device
IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer Contact
prashanth gali
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
7208902515
MDR Report Key7136003
MDR Text Key95459593
Report Number1723170-2017-05234
Device Sequence Number1
Product Code OXO
UDI-Device Identifier00643169353428
UDI-Public00643169353428
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K092564
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 12/20/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/20/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Radiologic Technologist
Device Model NumberBI-700-00028-120
Device Catalogue NumberBI70000028120
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/08/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received11/27/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/06/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-