Device is a combination product.(b)(4).A synergy ii us mr 2.50 x 24mm stent delivery system was returned for analysis.A visual examination of the crimped stent found no issues.There were no signs of damage, lifting or stretching of the stent struts.The crimped stent outer diameter was measured and was within maximum crimped stent profile measurement.The balloon cones were reviewed and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and tactile examination of the device found multiple kinks along the full catheter length.The hypotube was broken at approximately 787mm distal to the distal end of the strain relief.An examination of the shaft polymer extrusion found no issues.The bi-component bond showed no signs of damage.The tip was visually and microscopically examined and no issues were noted.No other issues were identified during the product analysis.The investigation conclusion is operational context as the product meets the design and manufacture specification but due to anatomical/ procedural factors encountered during the procedure, performance was limited.(b)(4).
|