(b)(4).Method: the complaint myairvo humidifier and a 900pt531 fph airvo pediatric heated breathing tube were received at fisher & paykel healthcare (fph) (b)(4) and were visually inspected and performance tested.An opt316 fph optiflow junior nasal cannula was also received and visually inspected.Multiple attempts were made to obtain further information from the customer regarding the incident, however no response was received.Results: the subject airvo and the 900pt531 heated breathing tube showed no signs of damage upon visual inspection.Visual inspection of the returned opt316 nasal cannula revealed the nasal prongs to be stained yellow, however no blockages were present.The 900pt531 heated breathing tube was connected to the airvo for performance testing and both were found to function within specification with no faults or defects.The airvo log showed no faults for the lifetime of the device.Additionally, the airvo log showed no usage time from the date of the incident.We note that the airvo log will record time used on any given day, provided that the airvo has been used for a minimum of 6 minutes.Conclusion: no faults were found with any of the returned devices and all were found to be operating within specification.The customer confirmed that the patient's condition stabilized once placed on mechanical ventilation.This suggests that the patient's worsened condition was not due to any malfunction of the airvo system, but rather may have been due to premature removal from mechanical ventilation.The airvo user manual states that the "airvo is for the treatment of spontaneously breathing patients who would benefit from receiving high-flow, warmed and humidified respiratory gases" and that "the unit is not intended for life support." it also states "use continuous oxygen monitoring on patients who would desaturate significantly in the event of disruption to their oxygen supply.".
|