MAUDE Adverse Event Report: BAXTER HEALTHCARE - AIBONITO CLEARLINK LUER ACTIVATED VALVE; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - AIBONITO CLEARLINK LUER ACTIVATED VALVE; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 2N8378
Device Problems Fluid/Blood Leak (1250); Hole In Material (1293)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/18/2017
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a clearlink catheter extension set had a hole near the blue clamp which resulted in a leak.The reporter stated that the tubing split shortly after the radiation technician injected computed tomography (ct) dye into the set during a chest scan.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
The actual device was not returned; however, a photograph of the device was received for evaluation.Visual inspection of the photograph revealed that the tube had split.The reported condition was verified.The cause of the condition was determined to be a use error.This set is not designed for use with power injection.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLEARLINK LUER ACTIVATED VALVE
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE - AIBONITO
aibonito
Manufacturer (Section G)
BAXTER HEALTHCARE - AIBONITO
rd 721 km 0 3 po box 1389
aibonito 00705
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key7137116
MDR Text Key95461522
Report Number1416980-2017-09922
Device Sequence Number1
Product Code FPA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112893
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number2N8378
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/29/2017
Initial Date FDA Received12/21/2017
Supplement Dates Manufacturer Received01/23/2018
Supplement Dates FDA Received01/25/2018
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Treatment
(B)(4)
-
-