Model Number H7493926224400
Device Problem Bent (1059)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/27/2017
Event Type  malfunction  
Manufacturer Narrative
Age at time of event: 18 years or older.Device is a combination product.(b)(4).The device was not returned for analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is operational context as the product meets the design and manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
 
Event Description
It was reported that stent damage occurred.Vascular access was obtained via the femoral artery.The 95% stenosed, 24mm in length, eccentric, de novo target lesion contained a greater than 45 and less than 90 degree bend and was located in the mildly tortuous, mildly calcified, and 3-4mm in diameter right coronary artery (rca).A 4.00 x 24 synergy stent was advanced but failed to cross the lesion.The device was removed and it was noted that the proximal part of the stent was deformed.The procedure was completed with another of the same device.No patient complications were reported and patient's status was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY¿
Type of Device
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
Manufacturer (Section G)
BOSTON SCIENTIFIC - GALWAY
Manufacturer Contact
sonali arangil
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key7137507
MDR Text Key95536111
Report Number2134265-2017-12232
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
SIMILAR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 11/27/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/09/2018
Device Model NumberH7493926224400
Device Catalogue Number39262-2440
Device Lot Number0020450402
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/27/2017
Initial Date FDA Received12/21/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/07/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
FC GUIDE WIRE; MACH1 SCR3.5 6F GUIDE CATHETER; RINATO GUIDE WIRE
-
-