• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM ONBOARD BATTERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC SYNCARDIA FREEDOM ONBOARD BATTERY Back to Search Results
Catalog Number 295025-001
Device Problems Detachment Of Device Component (1104); Device Damaged by Another Device (2915)
Patient Problems No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
Event Date 12/12/2017
Event Type  malfunction  
Manufacturer Narrative
This alleged failure mode poses a low risk to the patient because the battery was in the battery charger and not in use in the patient's freedom driver at the time of the customer-reported issue.The freedom onboard battery will be returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4) initial.
 
Event Description
The customer, a syncardia certified hospital, reported that the "prongs" on the bottom of the patient's freedom onboard battery pulled outward when removed from the battery charger.
 
Manufacturer Narrative
Physical inspection of the battery revealed it to exhibit connector / housing separation.In an attempt to reproduce the customer-reported condition (the 'prongs' on the bottom of freedom driver onboard battery pull outward when pulled out of the charger), the battery was installed and extracted three times in each bay of freedom battery charger returned with the battery.Following the installation / removal cycling of the battery, the battery terminals were examined and found to exhibit connector / housing separation, thereby duplicating and confirming the customer-reported issue.The root cause for the connector / housing separation is likely the result of a high number of insertion / extraction cycles the battery had accumulated during use in the field.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.(b)(4) follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM ONBOARD BATTERY
Type of Device
BATTERY
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer Contact
kerri hensley
1992 e. silverlake road
tucson, AZ 85713
5205451234
MDR Report Key7138728
MDR Text Key96126383
Report Number3003761017-2017-00258
Device Sequence Number1
Product Code LOZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Health Professional
Remedial Action Replace
Type of Report Initial,Followup
Report Date 10/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number295025-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/22/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/12/2017
Initial Date FDA Received12/21/2017
Supplement Dates Manufacturer Received12/12/2017
Supplement Dates FDA Received10/15/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/01/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age27 YR
-
-