The device was not returned to edwards for evaluation as it remained implanted.The device history record (dhr) could not be reviewed as a serial number was not provided.Aortic tears, or a tear of the aortic wall, may occur as a complication of cardiac surgery involving a diseased aortic root.Additional causes of aortic tears or a tear of the aortic wall may be related to the use or misuse of the edwards intuity valve.These aortic injuries are life threatening conditions that require urgent intervention with suture repair, pericardial patch repair or aortic root replacement.The ifu was reviewed and no inadequacies were identified with regards to warnings, contraindications, and the directions/conditions for the successful use of the device.No corrective or preventative actions apply to this case.A definitive root cause could not be determined.As there has been no confirmed design defect, manufacturing issue, or inadequacy in the labeling, ifu, or training leading to the complaint, edwards will continue to review of all reported events and perform trend analysis on a monthly basis.If action is required based on the determined control limits, appropriate investigation will be performed.No corrective or preventative actions are required at this time.
|
Edwards received notification that a bleeding into the left ventricle muscle near the aortic root was observed soon after the implantation of a 23mm valve.The surgeon added two pledged stitches on the muscle and then put glue and surgicel to stop the bleeding.Three sizers were tested 21mm (too loose), 23mm (comfort) and 25mm (too tight) so 23mm size was choose.The seating was easy and in very good place.As per medical opinion, the damage was not cause by the valve.The device remained implanted.The patient outcome was good and recovering well.
|