• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET FRANCE S.A.R.L. BIOMET BONE CEMENT R 40

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET FRANCE S.A.R.L. BIOMET BONE CEMENT R 40 Back to Search Results
Model Number N/A
Device Problem Unsealed Device Packaging (1444)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/02/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Report source- foreign.The event occurred in (b)(4).The product within this report is a combination product.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Not returned to manufacturer.
 
Event Description
It was reported the inner sterile packaging was compromised, allowing the cement powder to leak out.No further information is available at this time.
 
Manufacturer Narrative
(b)(4).Complaint samples was evaluated and the reported event was confirmed.Evaluation of returned products device found the supplier sealing were opened.The manufacturer sealing were intact.Device history record (dhr) was reviewed and no discrepancies were found.Investigation results concluded that the reported event was due to supplier packaging issue.Corrective action has been initiated to address reported issue.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOMET BONE CEMENT R 40
Type of Device
BONE CEMENT
Manufacturer (Section D)
BIOMET FRANCE S.A.R.L.
plateau de lautagne bp75
valence cedex 26903
FR  26903
Manufacturer (Section G)
BIOMET FRANCE S.A.R.L.
plateau de lautagne bp75
valence cedex 26903
FR   26903
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key7142667
MDR Text Key95915432
Report Number3006946279-2017-00294
Device Sequence Number1
Product Code MBB
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
PK150850
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date09/30/2021
Device Model NumberN/A
Device Catalogue Number4035890011
Device Lot NumberA641AJ1702
Other Device ID NumberSEE H10 NARRATIVE
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/04/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/27/2017
Initial Date FDA Received12/22/2017
Supplement Dates Manufacturer Received11/27/2017
Supplement Dates FDA Received02/02/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/03/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
-
-