• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. E1 44-36 STD HMRL BRNG; PROSTHESIS, SHOULDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. E1 44-36 STD HMRL BRNG; PROSTHESIS, SHOULDER Back to Search Results
Model Number N/A
Device Problem Difficult to Insert (1316)
Patient Problem No Information (3190)
Event Date 11/16/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported the bearing was unable to be implanted.No further information is available at this time.
 
Manufacturer Narrative
(b)(4).This follow up report is being submitted to relay additional information.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Concomitant medical products: comp rvs tray co 44mm cat: 115370 lot: 638790.Complaint sample was evaluated and the reported event was confirmed.Product was visually examined and dimensional analysis performed.All dimensions checked on the bearing are conforming.There was damage to the bearing in the extraction hole.Device history record was reviewed and no discrepancies relevant to the reported event were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
E1 44-36 STD HMRL BRNG
Type of Device
PROSTHESIS, SHOULDER
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key7142760
MDR Text Key95794774
Report Number0001825034-2017-11254
Device Sequence Number1
Product Code PAO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK113121
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 04/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/22/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date09/06/2022
Device Model NumberN/A
Device Catalogue NumberEP-115393
Device Lot Number024880
Other Device ID Number(01) 0 0880304 54315 7
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/08/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/17/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/06/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Age75 YR
-
-