No conclusion code available.Upon receipt, the device was interrogated with a programmer and the device image was retrieved.A longevity calculation was performed and the device was not within expected longevity performance.Test leads and loads were applied to the device and pacing lead impedance, high voltage impedance, pacing output, and sensing were tested and no anomalies were noted.The device was opened for further analysis and the current was measured and noted to be high.Fault isolation was performed, and the cause of the high current was traced to the hybrid.The reported event of premature battery depletion was confirmed.
|