Model Number CD3367-40Q
Device Problem Battery Problem (2885)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/11/2017
Event Type  Injury  
Event Description
Following the battery performance alert (bpa) advisory, a bpa was received by the clinician and the device was explanted.Further information was requested, but not yet available.
 
Manufacturer Narrative
No conclusion code available.Upon receipt, the device was interrogated with a programmer and the device image was retrieved.A longevity calculation was performed and the device was not within expected longevity performance.Test leads and loads were applied to the device and pacing lead impedance, high voltage impedance, pacing output, and sensing were tested and no anomalies were noted.The device was opened for further analysis and the current was measured and noted to be high.Fault isolation was performed, and the cause of the high current was traced to the hybrid.The reported event of premature battery depletion was confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA ICD
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key7143815
MDR Text Key95667301
Report Number2938836-2017-35360
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date 11/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/22/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2015
Device Model NumberCD3367-40Q
Device Lot Number4149496
Other Device ID Number05414734508339
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/04/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received11/05/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/03/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number2017865-08/28/17-001-C
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age85 YR
-
-