Model Number CD3251-40Q
Device Problem Premature Discharge of Battery (1057)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/03/2017
Event Type  Injury  
Event Description
It was reported that the device was explanted due to suspected premature battery depletion.There were no adverse consequences to the patient prior, during and post procedure.
 
Manufacturer Narrative
Premature battery depletion was confirmed by analysis.Bench testing on the device was performed, and no sources of high current were noted.In further analysis of the battery, lithium clusters were observed, but at the time of analysis the clusters did not appear to be in a location or size that would be sufficient to cause an internal short of the battery.As a result, the cause of the premature battery depletion could not be conclusively determined.However, from these analyses, in the absence of other root causes, it is probable that the premature battery depletion was caused by a lithium cluster induced short circuit.The device is included in the premature battery depletion with implantable cardioverter defibrillator advisory notice issued by st.Jude medical on 11 october 2016.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIFY QUADRA CRT-D, OUS MODEL, DF4COMP
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key7147024
MDR Text Key95789531
Report Number2938836-2017-35416
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup
Report Date 03/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2015
Device Model NumberCD3251-40Q
Device Catalogue NumberCD3251-40Q
Device Lot Number4180238
Other Device ID Number05414734504560
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/07/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/03/2017
Initial Date FDA Received12/26/2017
Supplement Dates Manufacturer Received03/15/2018
Supplement Dates FDA Received03/26/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/29/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-0116-2017
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age56 YR
-
-