Catalog Number 5MAXACE068KIT |
Device Problems
Fracture (1260); Device Damaged Prior to Use (2284); Out-Of-Box Failure (2311)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 11/29/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Results: the ace68 was kinked 57.5 cm from the hub.There were no visible fractures on the ace68.Conclusions: evaluation of the returned ace68 revealed a single kink along the mid-shaft of the catheter.If the catheter is retracted from the kit packaging prior to removing the top tray the catheter it may become pinned between the trays and kink.There were no visible fractures on the ace68.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
|
Event Description
|
During preparation for a thrombectomy procedure, the hospital staff noticed that a penumbra system ace 68 reperfusion catheter (ace68) was fractured mid shaft upon removal from the packaging.The damage to the ace68 was found prior to use, and therefore the ace68 was not used in the procedure.The procedure was completed using a new ace68.
|
|
Search Alerts/Recalls
|