• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AGA MEDICAL CORPORATION AMPLATZER PFO OCCLUDER; CARDIAC OCCLUSION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AGA MEDICAL CORPORATION AMPLATZER PFO OCCLUDER; CARDIAC OCCLUSION DEVICE Back to Search Results
Model Number 9-PFO-025
Device Problems Adverse Event Without Identified Device or Use Problem (2993); Unintended Movement (3026)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/30/2017
Event Type  Injury  
Manufacturer Narrative
An event of sizing issue was reported.The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
On (b)(6) 2017, an asd procedure was performed.Three devices (model: 9-asd-mf-025, lot: 6114176; model: 9-asd-010, lot: 4792862; model: 9-pfo-025, lot: 5934472) were attempted to be deployed, but all three were associated with sizing issues.The patient had to be taken to surgery for the last device (25 mm amplatzer pfo occluder) to be removed via a cut-down.Per report, there were no adverse consequences for the patient.The patient's anatomy did not accommodate the devices.
 
Event Description
Additional information received indicated the 25 mm amplatzer pfo occluder was surgically removed via a cut-down since it embolized.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER PFO OCCLUDER
Type of Device
CARDIAC OCCLUSION DEVICE
Manufacturer (Section D)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key7148963
MDR Text Key95858883
Report Number2135147-2017-00194
Device Sequence Number1
Product Code MLV
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P120021
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date03/31/2022
Device Model Number9-PFO-025
Device Catalogue Number9-PFO-025
Device Lot Number5934472
Other Device ID Number00811806011288
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/30/2017
Initial Date FDA Received12/27/2017
Supplement Dates Manufacturer Received12/28/2017
Supplement Dates FDA Received01/15/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/05/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age75 YR
-
-