The reported events of a capture issue and the connector being too loose were not confirmed.As received, a complete lead was returned in one piece for analysis.Visual inspection of the lead did not find any anomalies at the connector region or throughout lead body.Dimensional analysis revealed the diameters of the is-4 connector pin, connector ring electrodes, and the connector seal zones to be within specification.An insertion test was performed with a test device and no anomalies were noted.Electrical tests that could be performed on the returned portion of the lead did not find any conductor fractures or internal shorts.
|