The device history record review confirms that the device met all material, assembly and performance specifications.Analysis of the device revealed that the coil delivery wire was broken at the main coil junction area.In addition, the main coil was found to be kinked and stretched likely due to procedural factors.Functional testing could not be performed due to the condition of the returned device and as the delivery wire was not returned.It is probable that the break at the main coil junction area is due to handling of the device or portion of the device during the clinical procedure, upon removal of the device from the packaging, or preparation of the device prior to use, thus limiting the performance of the device.Therefore an assignable cause of handling damage will be assigned to the analyzed coil delivery wire broken.
|