Model Number MO4S
Device Problems Burst Container or Vessel (1074); Leak/Splash (1354); Melted (1385); Overheating of Device (1437)
Patient Problem Burn, Thermal (2530)
Event Date 12/27/2017
Event Type  Injury  
Event Description
It's about 1 am and the malem bed-wetting alarm that my daughter has strapped on has burnt her.The alarm is extremely hot and the back door that snaps in and holds batteries in place has melted from the heat.It is unfortunate that this happened.Additionally, the heat has caused the two aaa batteries inside the alarm part to burst and leak onto my daughters clothing and seep in to her neck and chest.She has had a reaction to the battery liquid and has been slightly burnt from both battery liquid and the hot bed-wetting alarm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MALEM BED-WETTING ALARM
Type of Device
MALEM BED-WETTING ALARM
Manufacturer (Section D)
MALEM MEDICAL (UK)
MDR Report Key7156793
MDR Text Key96225431
Report NumberMW5074309
Device Sequence Number1
Product Code KPN
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Reporter Occupation Other
Type of Report Initial
Report Date 12/28/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Device Operator Lay User/Patient
Device Model NumberMO4S
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received12/29/2017
Was Device Evaluated by Manufacturer? No Information
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age5 YR
Patient Weight22
-
-