Model Number N/A |
Device Problems
Migration or Expulsion of Device (1395); Insufficient Information (3190)
|
Patient Problems
Bone Fracture(s) (1870); No Information (3190); No Code Available (3191)
|
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).Medical devices: unknown, unknown ulnar component, unknown.Customer has indicated that the product will not be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001822565 - 2017 - 08618.
|
|
Event Description
|
It was reported that the patient underwent an initial right elbow arthroplasty on an unknown date.Subsequently, the patient has been indicated for a revision due to unknown reasons.No further information has been provided.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.Additional: updated.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Manufacturer Narrative
|
(b)(4).This follow-up report is being submitted to relay additional information.Date of event - unknown date in (b)(6) 2017.Date of implant - unknown date in 2007.Date of explant - unknown date in (b)(6) 2017.
|
|
Event Description
|
It was reported that the patient underwent an initial right elbow arthroplasty and was revised due to loosening of the humeral prosthesis.No further information has been provided.
|
|
Manufacturer Narrative
|
Reported event was confirmed by review of x-rays provided and notes provided by rep.X-ray reviewer stated "significant radiolucency is identified at the bone cement interface of all of the images of the humeral component which progresses until fracture of the humerus.Some radiolucency is also identified at the bone cement interface of the ulnar component.Device history record (dhr) review was unable to be performed as the lot number of the device involved in the event is unknown.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|