AGA MEDICAL CORPORATION AMPLATZER MUSCULAR VSD OCCLUDER; CARDIAC OCCLUSION DEVICE Back to Search Results
Model Number 9-VSD-MUSC-012
Device Problem Difficult To Position (1467)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/26/2017
Event Type  malfunction  
Manufacturer Narrative
Product investigation: the reported event of a failure to retract the amplatzer muscular vsd occluder into the 7f delivery system sheath could not be confirmed.The returned occluder met visual and functional specifications when analyzed at abbott.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported event remains unknown.
 
Event Description
On (b)(6) 2017, while pulling back a 7f amplatzer torqvue delivery system 45 (dtv45), the amplatzer muscular vsd occluder was in the incorrect position.The occluder failed to retract into the dtv.The dtv45 and occluder were removed from the patient and the sheath appeared damaged at the tip.On a later date, another amplatzer muscular vsd occluder was used and procedure was successfully completed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER MUSCULAR VSD OCCLUDER
Type of Device
CARDIAC OCCLUSION DEVICE
Manufacturer (Section D)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key7161707
MDR Text Key96728086
Report Number2135147-2017-00179
Device Sequence Number1
Product Code MLV
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P040040
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Reporter Occupation Health Professional
Type of Report Initial
Report Date 01/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2022
Device Model Number9-VSD-MUSC-012
Device Catalogue Number9-VSD-MUSC-012
Device Lot Number5888293
Other Device ID Number00811806011820
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/05/2017
Initial Date FDA Received01/03/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/06/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Other;
-
-