Catalog Number 06437281190
Device Problem High Test Results (2457)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/18/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).This event occurred in (b)(6).
 
Event Description
The customer received a questionable high elecsys ft4 ii assay result for one patient with hypothyroidism.Previously, the patient was diagnosed with hyperthyroidism (grave's disease) as indicated by high ft3 and ft4 levels.After treatment, the patient symptoms appeared to be hypothyroid, but the ft4 result was still high.The physician thought the ft4 result should not be high and suspected an interference.On (b)(6) 2017, the ft4 result from a cobas e 411 immunoassay analyzer was 2.49 ng/dl.On (b)(6) 2017, the sample was repeated in another laboratory using cmia-architect and the result was 1.91 ng/dl.The result was reported outside of the laboratory.There was no allegation of an adverse event.The serial number of the cobas e 411 immunoassay analyzer was requested but was not provided.
 
Manufacturer Narrative
As no sample material from the patient was available for further investigation, a specific root cause could not be determined.From the information provided, a general reagent issue could most likely be excluded.Ft4 results generated with different types of analyzers, can be different due to the overall setups of the assays, the antibodies used, and differences in the standardization materials/methodologies used.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FT4, FREE THYROXINE
Type of Device
RADIOIMMUNOASSAY, FREE THYROXINE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key7163602
MDR Text Key97374021
Report Number1823260-2018-00064
Device Sequence Number1
Product Code CEC
Combination Product (y/n)N
Reporter Country CodeTH
PMA/PMN Number
K961489
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 01/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number06437281190
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/22/2017
Initial Date FDA Received01/04/2018
Supplement Dates Manufacturer Received12/22/2017
Supplement Dates FDA Received01/29/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-