Catalog Number 06437281190 |
Device Problem
High Test Results (2457)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 12/18/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).This event occurred in (b)(6).
|
|
Event Description
|
The customer received a questionable high elecsys ft4 ii assay result for one patient with hypothyroidism.Previously, the patient was diagnosed with hyperthyroidism (grave's disease) as indicated by high ft3 and ft4 levels.After treatment, the patient symptoms appeared to be hypothyroid, but the ft4 result was still high.The physician thought the ft4 result should not be high and suspected an interference.On (b)(6) 2017, the ft4 result from a cobas e 411 immunoassay analyzer was 2.49 ng/dl.On (b)(6) 2017, the sample was repeated in another laboratory using cmia-architect and the result was 1.91 ng/dl.The result was reported outside of the laboratory.There was no allegation of an adverse event.The serial number of the cobas e 411 immunoassay analyzer was requested but was not provided.
|
|
Manufacturer Narrative
|
As no sample material from the patient was available for further investigation, a specific root cause could not be determined.From the information provided, a general reagent issue could most likely be excluded.Ft4 results generated with different types of analyzers, can be different due to the overall setups of the assays, the antibodies used, and differences in the standardization materials/methodologies used.
|
|