MAUDE Adverse Event Report: ZIMMER BIOMET SPINE INC. CANNULATED POLYAXIAL SCREW 6.5MM X 50MM; PATHFINDER NXT MINIMALLY INVASIVE PEDICLE SCREW SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET SPINE INC. CANNULATED POLYAXIAL SCREW 6.5MM X 50MM; PATHFINDER NXT MINIMALLY INVASIVE PEDICLE SCREW SYSTEM Back to Search Results
Catalog Number 3505-6550
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The returned screw was evaluated.The screw shaft was damaged in a manner consistent with tightening and subsequent loosening, which allows the tulip to disassemble from the screw shaft.Additionally, the threads within the tulip were damaged in a manner consistent with cross-threading the closure top within the tulip during assembly and tightening.A review of the manufacturing records did not identify any issues which would have contributed to this event.
 
Event Description
It was initially reported that a pedicle screw "didn't feel right" during surgery.The screw was removed and replaced with an alternative screw to complete the procedure.There were no reports of patient injury associated with this event.Device evaluation by zimmer biomet personnel found that the screw had disassembled and there was thread damage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CANNULATED POLYAXIAL SCREW 6.5MM X 50MM
Type of Device
PATHFINDER NXT MINIMALLY INVASIVE PEDICLE SCREW SYSTEM
Manufacturer (Section D)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer (Section G)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
na
westminster CO 80021
Manufacturer Contact
geoffrey gannon
10225 westmoor dr.
na
westminster, CO 80021
3034437500
MDR Report Key7165738
MDR Text Key96483954
Report Number3012447612-2018-00017
Device Sequence Number1
Product Code NKB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK132884
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 01/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number3505-6550
Device Lot Number85LX
Other Device ID Number(01)00889024335714(10)85LX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/13/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/03/2018
Initial Date FDA Received01/04/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/14/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-