• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HILL-ROM BATESVILLE TOTAL CARE FRAME; A/C POWERED ADJUSTABLE HOSPITAL BED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HILL-ROM BATESVILLE TOTAL CARE FRAME; A/C POWERED ADJUSTABLE HOSPITAL BED Back to Search Results
Model Number P1900G006203
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/22/2017
Event Type  malfunction  
Manufacturer Narrative
The hill-rom technician found the power control board needed to be replaced.The power control board houses the external speaker, which is the source of the brake not set alarm.Per the hill-rom service manual the totalcare bed system requires an effective maintenance program.We recommend that you perform a semi-annual preventative maintenance.Preventative maintenance will minimize downtime due to excessive wear.Test the brake casters to determine if the bed moves when you activate the brake, replace or adjust when necessary.Check the tires for cuts, wear, tread life, etc.A search of the hill-rom maintenance records showed hill-rom performed preventative maintenance on this bed in 2017.It is unknown if the facility performed any other preventative maintenance on this bed.The technician replaced the power control board to resolve the issue.Based on this information, no further action is required.
 
Event Description
Hill-rom received a report from a hill-rom technician stating the bed alarm was not sounding.The bed was located at the account.There was no patient/user injury reported.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOTAL CARE FRAME
Type of Device
A/C POWERED ADJUSTABLE HOSPITAL BED
Manufacturer (Section D)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
tony werner
1069 state route 46 east
batesville, IN 47006
8129312359
MDR Report Key7167517
MDR Text Key96592188
Report Number1824206-2018-00011
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K962942
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 12/22/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/05/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other Caregivers
Device Model NumberP1900G006203
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received12/22/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/11/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-