MAUDE Adverse Event Report: ADVANCED BIONICS, LLC HIRES 90K¿ IMPLANT; COCHLEAR IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ADVANCED BIONICS, LLC HIRES 90K¿ IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number CI-1400-01
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Code Available (3191)
Event Date 12/08/2017
Event Type  malfunction  
Event Description
The recipient is reportedly experiencing loss of lock.External equipment was exchanged and programming adjustments were made, however, the issue did not resolve.The recipient was a non-user of the device.Revision surgery will be scheduled.
 
Manufacturer Narrative
The recipient's device was explanted.The recipient was reimplanted with another cochlear device.
 
Manufacturer Narrative
The external visual inspection revealed that the array was severed prior to receipt.This is believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock could not be obtained at any spacing.The no lock condition prevented some of the electrical tests from being performed.The device passed an electrical test performed.The device failed the residual gas analysis test.The internal visual inspection noted silver migration across some electrical components.This device had moisture that exceeded the residual gas analysis test limit.The source of the problem was a feedthru hermeticity issue from one feedthru vendor.A corrective action was implemented.Feedthru assemblies from this vendor are no longer used.This is the final report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES 90K¿ IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
Manufacturer Contact
lacey mcdonald
28515 westinghouse place
valencia, CA 91355
6613621741
MDR Report Key7169315
MDR Text Key96729553
Report Number3006556115-2017-00715
Device Sequence Number1
Product Code MCM
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Audiologist
Type of Report Initial,Followup,Followup
Report Date 12/08/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date04/30/2007
Device Model NumberCI-1400-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/03/2018
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received01/05/2018
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received03/21/2018
04/26/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/30/2005
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age26 YR
-
-