Results: the ace68 was fractured approximately 70.0 cm from the proximal end.Conclusions: the complaint has been evaluated.The complaint indicated that during preparation for a thrombectomy procedure, the hospital technologist noticed that an ace68 was snapped in half upon removal from the packaging.This damage may have occurred due to forceful handling at extreme angles during removal from the packaging tray.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
During preparation for a thrombectomy procedure, the hospital technologist noticed that a penumbra system ace 68 reperfusion catheter (ace68) was snapped in half upon removal from the packaging of the penumbra system ace 68 hi-flow kit (kit).The damaged ace68 was found prior to use.Therefore, the ace68 was not used in the procedure.The procedure was completed using another ace68.
|