Model Number ROTAFLOW CONSOLE
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/17/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).A follow-up medwatch will be submitted when additional information becomes available.The field service technician (fst) was sent for further investigation.After a longtime test and a lot of restarts with good results, the fst were not able to reproduce the error.The device works as intended.The device is back in clinical use.Since no parts were defected and therefore no parts has been replaced, no further investigation is possible.Thus the failure could not be confirmed.The data is also being handled through a designated maquet cardiopulmonary trending and applicable investigation process.Due to this no further investigation initiations will be completed at this time.Since the reported failure did not contribute to a death or serious injury no corrective action is needed.In addition at this time it cannot be concluded that this is a systemic error.No corrective action is needed.
 
Event Description
Double assist.After about 20 hours of use both rotaflow get alarm "sig".They add some more ultrasonic coupling paste to the flow sensor and one rotaflow worked again.On the other rotaflow unit they add some more paste without good result.Turned of the rotaflow and started it again but did not get it to work.Alarm and error.Didn't saw what the message was.Used the emergency drive and changed the unit to a other rotaflow.No known consequences to the patient.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTAFLOW CENTRIFUGAL PUMP SYSTEM
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY AG
rastatt
GM 
Manufacturer (Section G)
BERND RAKOW
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt
GM  
Manufacturer Contact
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt 
4972229321
MDR Report Key7175060
MDR Text Key96839956
Report Number8010762-2018-00012
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeSW
PMA/PMN Number
K991864
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/09/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberROTAFLOW CONSOLE
Device Catalogue NumberJH10.51696
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/18/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/01/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-