Model Number MICROBLENDER
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Date 12/20/2017
Event Type  malfunction  
Manufacturer Narrative
Vyaire file identification: (b)(4).Any additional information received from the customer will be included in a follow-up report.Results of investigation: the vyaire failure analysis laboratory received the suspect microblender assembly for investigation.An evaluation of the component could be confirmed and duplicated.The dry o-rings on the poppet assembly was causing the poppet to stick and not alarm.The alarm adjustments were also found to be out of calibration.
 
Event Description
The customer reported while using the p/n 9320 low flow air/oxygen microblender; the device's alarm is not working.The customer reported the device is due for overhaul and repairs.The customer reported there is no patient involvement associated with the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
P/N 9320 LOW FLOW AIR/OXYGEN MICROBLENDER
Type of Device
MIXER, BREATHING GASES, ANESTHESIA INHALATION
Manufacturer (Section D)
VYAIRE MEDICAL, INC
22745 savi ranch parkway
yorba linda CA 92887
Manufacturer (Section G)
VYAIRE MEDICAL, INC
1100 bird center dr.
palm springs CA 92262
Manufacturer Contact
kristin graf
22745 savi ranch parkway
yorba linda, CA 92887
MDR Report Key7175631
MDR Text Key96773718
Report Number2021710-2018-07271
Device Sequence Number1
Product Code BZR
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K883038
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 01/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/09/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMICROBLENDER
Device Catalogue Number10065
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/28/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/20/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/01/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-