Catalog Number 07.02010.001
Device Problems Migration or Expulsion of Device (1395); Use of Device Problem (1670); Device Operates Differently Than Expected (2913)
Patient Problems No Information (3190); No Code Available (3191)
Event Date 08/30/2016
Event Type  Injury  
Manufacturer Narrative
Without a product return, no product evaluation is able to be conducted.The lot number is unknown; therefore the device history records are unable to be reviewed.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
 
Event Description
It was reported that a plug popped off an iliac pedicle screw during surgery.Additional information has been requested but not made available.
 
Manufacturer Narrative
The surgeon provided notes that states he felt this event was caused by user error, and was not the result of a device malfunction.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
 
Event Description
It was reported that a few months after surgery, the patient bent over and felt a pop in her back, and subsequently had discomfort.An x-ray taken after this showed that the closure top had backed out of the pedicle screw head.A revision surgery was performed to remove and replace the migrated closure top.There have been no further patient events or issues.
 
Manufacturer Narrative
The device was not returned.However, a photo of an x-ray was provided and used to confirm the closure top has migrated from its original location.The lot number is unknown; therefore the device history records are unable to be reviewed.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLOSURE TOP
Type of Device
VITALITY SPINAL FIXATION SYSTEM
Manufacturer (Section D)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
westminster CO 80021
Manufacturer (Section G)
ZIMMER BIOMET SPINE INC.
10225 westmoor dr.
na
westminster CO 80021
Manufacturer Contact
geoffrey gannon
10225 westmoor dr.
na
westminster, CO 80021
3034437500
MDR Report Key7180319
MDR Text Key96907954
Report Number3012447612-2018-00044
Device Sequence Number1
Product Code NKB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK150896
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 08/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/10/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number07.02010.001
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/05/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age52 YR
Patient Weight59
-
-