Product analysis: upon receipt at medtronic¿s quality laboratory, visual examination revealed the valve was intact with no evidence of visible damage.Both leaflets were received in the closed position and were intact with no evidence of damage such as cracks and/or surface anomalies.Both inflow and outflow valve hinge mechanisms were intact.The inflow and outflow orifices were intact with no evidence of damage.Using a blue actuator to test leaflet movement, the leaflets moved without difficulty.Conclusion: the device history record was reviewed and showed that this product met all manufacturing specifications for product released for distribution.No issues were identified that would have impacted this event.Based on the analysis, the cause of the event cannot be determined as no leaflet motion issue was noted.The valve annulus diameter was verified and met the specification as a 29mm aortic valve.Based on the received information, the leaflet motion issue could be due to sizing.If information is provided in the future, a supplemental report will be issued.
|